Cautions
b Ensure that the connector and most of the cable remain out of the fluid. You may submerge 2 inches
(5 cm) of the cable proximal to the transducer scanhead.
Connector
Do not submerge
Figure 1
Transducer component names
12 Rinse the transducer 3 separate times using the following procedure:
a Rinse the transducer in clean, running water according to the disinfectant manufacturer's instructions
(at least 1 minute).
b Ensure that the connector and at least 12-18 inches (31-46 cm) of the cable from the connector remain
dry.
c Repeat this step until the transducer has been rinsed 3 separate times.
13 Dry the transducer with a sterile lint-free cloth.
14 Dispose of the disinfectant according to the manufacturer's guidelines.
14
Do not soak the transducer longer than recommended by the chemical
manufacturer.
Do not immerse the transducer connector in any disinfectant solution.
Use only FUJIFILM SonoSite recommended cleaners and disinfectants. Using a
non- recommended disinfecting solution or incorrect solution strength can
damage or discolor the transducer and void the warranty.
Cable
Maximum
5 cm
(2 inches)
Scan head
Cleaning and disinfecting