2.7.
Scale Capacity and Increment Display Options
2.7.1.
Metrology Line
In order to meet metrological requirements from different regions of the world, a metrological line
that shows capacity and increment information or Max/Min/e information on the IND570 display
can be enabled within terminal configuration (Figure 2-63). When enabled, this line is always
shown except when in the setup mode. If multiple ranges are selected, this line will scroll through
each of the available capacities and ranges.
Refer to the Scale and Terminal sections in Chapter 3, Configuration, for details on enabling the
Metrology Line in IND570. In cases where the metrology line of data is not accepted by an agency,
it can be disabled and a label with the required information may be installed near the display.
Metrology Line
2.7.2.
Capacity Label Instructions
To further ensure that scale capacity and increment display requirements can be met during
installation, a blue capacity label is included with the terminal. This label can be completed and
adhered to the front of the terminal overlay as shown in Figure 2-65. To prevent any potential
conflicts, when using the capacity label, the metrology line on the display should be disabled in
setup.
The capacity label (shown in Figure 2-64) provides space for the Max, min, and e information for
each range or interval for which the scale is programmed. If only one or two ranges are used, the
unused portion of the label may be cut off with scissors. Written information must be legible and a
minimum of 2mm or 0.08 in. in height. A permanent marker should be used for this information.
30205319
|
04
|
08/2016
Figure 2-63: Metrology Line on IND570 Display
METTLER TOLEDO IND570 Weighing Terminal Installation Manual
2-43