1.3.
Environmental Protection
When an approved version IND570xx is installed in an area classified as Division 2 or Zone 2/22,
special AC wiring requirements must be met. Refer to document 30205321, IND570xx Division 2,
Zone 2/22 Installation Guide.
30205319
|
04
|
08/2016
IND570 Specifications
Hazardous Area
US and Canada
CL I, DIV 2, GP CD; CL II, DIV2, GP FG; CL III; CL I Zn
2 AEx / Ex ic nA [ic] IIB T5; Zn 22 AEx tc IIIC T85°C -
10°C ≤ Ta ≤ +40°C Temperature ID = T5 (100°C)
Europe
II 3 G Ex ic nA [ic] IIB T5 Gc -10°C ≤ Ta ≤ +40°C
II 3 D Ex tc IIIC T85°C Dc -10°C ≤ Ta ≤ +40°C
Global
Ex ic nA [ic] IIB T5 Gc -10°C ≤ Ta ≤ +40°C
Ex tc IIIC T85°C Dc -10°C ≤ Ta ≤ +40°C
VERSIONS OF THE IND570 TERMINAL THAT ARE NOT FACTORY-LABELED AS DIVISION 2 OR
ZONE 2/22 MUST NOT BE INSTALLED INTO A DIVISION 2 OR ZONE 2/22 ENVIRONMENT.
THE IND570xx TERMINAL IS NOT INTRINSICALLY SAFE LEVEL "a" OR "b"! DO NOT USE WITHIN
AREAS CLASSIFIED AS HAZARDOUS DIVISION 1 OR ZONE 0/1/20/21 BECAUSE OF
COMBUSTIBLE OR EXPLOSIVE ATMOSPHERES. FAILURE TO COMPLY WITH THIS WARNING
COULD RESULT IN BODILY HARM AND/OR PROPERTY DAMAGE.
METTLER TOLEDO IND570 Weighing Terminal Installation Manual
WARNING
WARNING
IP65
IP65
IP65
1-7