IT Network
Functions
This device can be connected to an IT network to perform the following functions:
Storing the examination data (static images, clips) acquired by this device in a Picture Archiving and
Communication System (PACS) by DICOM communication.
Querying examination orders from the Modality Worklist (MWL) server by DICOM communication and
starting them.
Uploading images to the SonoSite Patient Data Archival Software (PDAS) or SiteLink Image Manager.
Setting the time of this device by interrogating the network time service.
Communicating procedure status via the Modality Performed Procedure Step (MPPS) service.
Requesting transfer of responsibility for image ownership to another system via the Storage Commitment
service.
Network for connecting the device
To ensure safety, use an IT network that is isolated from the external environment by a firewall.
Specifications for the connection
Hardware specification
802.11 b/g/n
Ethernet 100BASE-TX/10BASE-T using RJ45 port with patch cable
Software specifications
This device is connected to PACS and MWL by the DICOM standard. Refer to the DICOM Conformance
Statement of this device for details.
When available, this device connects to the network time server at startup.
The system conforms to the DICOM standard as specified in the SonoSite Edge II, SonoSite SII DICOM
Conformance Statement, available at www.sonosite.com. This statement provides information about the
purpose, characteristics, configuration, and specifications of the network connections supported by the
system.
Security
The port for DICOM communication (specified by the user in the system settings; typically port 104, 2762,
or 11112) is used for outgoing communication to the network.
Anti-virus software is not installed on this device.
6
IT Network