6.
Reassemble the PureWick™ Urine Collection System according to the Initial Setup
instructions when the system is ready to be reused.
When the PureWick™ Urine Collection System is not in use, unplug the power cord from
its outlet. Make sure the unit is cleaned and disinfected prior to storage. Do not store
when parts are wet or damp.
Cleaning and Maintenance
Inspect the PureWick™ Urine Collection System and accessories for damage or wear
prior to use and replace as necessary.
DO NOT ATTEMPT TO OPEN OR DISMANTLE THE PUREWICK™ URINE
COLLECTION SYSTEM. THIS MAY AFFECT PERFORMANCE, CREATE A POTENTIAL
SAFETY HAZARD, CAUSE PERSONAL INJURY, AND WILL VOID THE WARRANTY.
ALWAYS TURN OFF THE DEVICE AND DISCONNECT FROM POWER SOURCE PRIOR
TO CLEANING.
THE PRIVACY COVER COMPONENTS ARE SINGLE USE ONLY AND SHOULD NOT BE
CLEANED OR DISINFECTED.
Collection Canister
and Lid
Frequency:
P
Daily
P
Single Patient
USE CASE
Use Only
Point of Use Cleaning: After the completion of each use, the collection canister, canister
lid, collector tubing, and pump tubing should be rinsed with cool tap water to remove
any residual debris or dirt. The PureWick™ Urine Collection System base should be wiped
with a clean low lint cloth dampened with cool tap water.
Manual Cleaning Instructions for Single User (Home Care):
The collection canister, canister lid, collector tubing, pump tubing, and PureWick™ Urine
Collection System base should be cleaned and disinfected at the time of each use, or at
a minimum, daily. The power cord should be cleaned and disinfected at the time of each
use, or at a minimum daily.
Note: Gloves should be worn when handling soiled or dirty accessories.
Canister and Canister Lid
• Initial Rinse: Rinse the canister and lid thoroughly with cool tap water. While rinsing,
remove any excess dirt by wiping the canister and lid with disposable low lint wipes.
Collector Tubing
with elbow connector
P
Clean
Clean
P
Disinfect
Disinfect
Single Patient
Use Only
PureWick™ Urine
Collection System base
P
Clean
P
Disinfect
Reusable
5
Pump Tubing
Power Cord
P
P
Clean
Clean
P
P
Disinfect
Disinfect
Single Patient
Reusable
Use Only