PureWick™ Urine Collection System Base with power cord
• Initial Wipe: Wipe the PureWick
thoroughly with a clean low lint cloth dampened with cool tap water. Remove any
excess dirt by wiping the device with disposable low lint wipes.
• Cleaning: Prepare a soapy solution by mixing 1 teaspoon (approximately 5 mL)
of dish soap with 1 gallon (approximately 4 L) of cool tap water. Completely wet a
clean low lint cloth with the solution and wipe down the PureWick
System base and power cord. Allow the cleaning solution to remain on the device
and power cord for a minimum of ten (10) minutes, ensuring that the device and
power cord remains wet for the ten (10) minutes. Using a soft brush (e.g. toothbrush),
brush all accessible areas of the PureWick
cord for a minimum of one (1) minute to remove any visible signs of debris or dirt.
• Rinse: Using a new clean low lint cloth, wipe the PureWick
System base and power cord thoroughly with cool tap water to remove the
soap cleaning solution. Perform this step until there is no visible sign of the soap
cleaning solution.
• Visual inspection: Inspect the PureWick
power cord to ensure that all debris and dirt have been removed. If there is any
sign of debris or dirt, repeat the steps above until there is no sign of debris or dirt
remaining on the PureWick
• Disinfection: Use 70% isopropyl alcohol (IPA) to completely wet a clean low lint
cloth and wipe down the PureWick
cord. Allow the disinfecting solution to remain on the device and power cord for
a minimum of ten (10) minutes, ensuring the unit is visibly wet with IPA for ten
(10) minutes.
• Rinse: Using a new clean low lint cloth, wipe the PureWick
System base and power cord thoroughly with cool tap water to remove the
disinfecting solution.
• Drying: Dry the PureWick
low lint towel or cloth.
Manual Cleaning Instructions for Hospitals and Long-Term Care Facilities:
Contamination should be assumed prior to use with a new user; therefore, the PureWick™
Urine Collection System base and power cord must be fully cleaned with a disinfectant
before reuse. The collection canister and tubing are not suitable for reuse in a multiple user
environment. A new collection canister and new tubing must be used for each new user.
Canister and Canister Lid (Single User)
• Initial Rinse: Rinse the canister and lid thoroughly with cool tap water. While rinsing,
remove any excess dirt by wiping the canister and lid with disposable low lint wipes.
• Cleaning: Prepare enzymatic cleaner solution (e.g. Enzol Enzymatic Cleaner or
equivalent) following the manufacturer's recommendation of 1 oz. (approximately
29 mL) per gallon (approximately 4 L) of warm tap water (ratio and water
temperature per manufacturer's recommendation if applicable).
Urine Collection System base and power cord
TM
TM
TM
Urine Collection System base and power cord.
TM
Urine Collection System base and power
TM
Urine Collection System base and power cord with a clean
TM
7
Urine Collection System base and power
TM
Urine Collection System base and
TM
Urine Collection
TM
Urine Collection
Urine Collection