4.3 Preparation

If a pressure gauge has been used:
• Detach the pressure gauge connector line from the connect-
ing element.
• Detach the connection tubing from the pressure gauge.
• Detach the tubing adapter from the pressure gauge con-
nector line and the connection tubing.
General:
• If necessary, disconnect the PARI PEP S system from the
nebuliser.
• Dismantle the PARI PEP S system into its individual parts.
• Open the cap on the connecting element.

4.4 Cleaning

Precleaning
All individual parts must be precleaned immediately after use.
EQUIPMENT:
– Drinking water temperature of about 15 °C
PROCEDURE:
• Rinse all parts used for 2 minutes in running drinking water.
Manual cleaning
EQUIPMENT:
– Drinking water with a temperature of at least 40 °C
– Standard commercial washing-up liquid
– Receptacle with at least 3 l capacity
3) Validiert mit Palmolive
– 20 –
®
.