5 With the required reference input applied to the correct terminals, press Hold/
6 During calibration, the primary display and bar-graph will indicate the
7 Repeat step 4 and step 5 until all the adjustment items for the particular
8 Select another function to be calibrated. Repeat step 4 to step 7. For a rotary
9 After calibrating all the functions, switch off the instrument and then switch it
You may also refer to
Keysight U1211A, U1212A, and U1213A User's and Service Guide
– If the required reference input is "OPEN", just leave the terminals open.
– If the required reference input is a voltage, current, resistance, capacitance,
or temperature value, set up the Fluke 5520A calibrator (or another device
with equivalent standard of accuracy) to provide the necessary input.
Max Min to start the present adjustment item.
uncalibrated reading, and the calibration indicator, "CAL", will appear on the
secondary display. If the reading is within the acceptable range, the word
"PASS" will be shown momentarily, and then the instrument will proceed to
the next adjustment item. If the reading is out of the acceptable range, it will
remain at the present adjustment item after showing the error code for 3
seconds. In this case, you need to check whether the correct reference input
has been applied. Refer to
page 115 for the meaning of the error codes.
function have been completed.
switch position that hosts more than one function, for example,
Shift/Peak to go to the next function.
on again. The instrument will be back to normal measurement mode.
"Typical adjustment process flow"
Performance Tests and Calibration
"Error codes and their respective meanings"
on
, press
on page 108.
6
107