European Certifications:
CSA International (Canadian Standards Association)
Protections Methods:
Class 2258 02 – Process Control Equipment – For Hazardous Locations (CSA 1078546)
Class I, Division 1, Groups B, C and D
Class II, Division 1, Groups E, F and G
Class III, Division 1
Class I, Division 2, Groups A, B, C and D
Class 2258 04 – Process Control Equipment – Intrinsically Safe, Entity – For Hazardous
Locations (CSA 1078546)
Exia - Class I, Division 1, Groups A, B, C and D
Exia - Class II, Division 1, Groups E, F and G
Exia - Class III, Division 1
Intrinsically safe with entity parameters: Vmax = 28 V Imax = 110 mA Ci = 5 nF Li = 12 uH,
when connected through CSA Certified Safety Barriers as per Smar Installation Drawing:
102A-0833.
Environmental Protection (CSA 1078546)
Options: Type 4X or Type 4
Special conditions for safe use:
Temperature Class T3C
Maximum Ambient Temperature: 40ºC
Drawings for manuals:
Label Plate: 102A-1392 and 102A-1423
Installation Drawing: 102A-0833
NEMKO (Norges Elektriske MaterielKontroll)
Protections Methods:
Explosion Proof (Nemko 00ATEX305)
Group II, Category 2 G, Ex d, Group IIC, Temperature Class T6
Maximum Ambient Temperature: 40ºC
Environmental Protection (Nemko 00ATEX305)
Options: IP66 W or IP66
Drawings for manuals:
Label Plate: 102A-1412 and 102A1494
EXAM (BBG Prüf - und Zertifizier GmbH)
Protections Methods:
Intrinsic Safety (DMT 01ATEX E 012)
Group II, Category 2 G, Ex d [ia], Group IIC, Temperature Class T6
Entity Parameters: Ui = 28 Vdc Ii = 93 mA Pi = 700 mW Ci ≤ 5 nF Li = neg
Ambient Temperature: -20ºC ≤ Ta ≤ + 40ºC
Drawings for manuals:
Label Plate: 102A-1412 and 102A1494
Informaciones sobre Certificaciones
A.3