Thank you for selecting the Blood Pressure Monitor with Alarm Clock. The monitor features
blood pressure measurement, pulse rate measurement, and record memory. The device is
designed for up to 2 users.
Readings taken by the Upper Arm Blood Pressure Monitor are equivalent to those obtained by
a trained observer using the cuff and stethoscope auscultation method.
This manual contains important safety and care information, and provides step by step
instructions for using the product. Please read the manual thoroughly before using the
product. Please keep this manual for future reference.
• This device is intended only for adult use in homes.
• This device is intended for non-invasive measuring and monitoring of arterial blood
pressure. It is not intended for use on extremities other than arm or for functions other than
obtaining a blood pressure measurement.
• If you experience discomfort during a measurement, such as pain in the arm or other
complaints, press the START/STOP button to release the air immediately from the cuff.
Loosen the cuff and remove it from your arm.
• On the rare occasion of a fault causing the cuff to remain fully inflated during measurement,
open the cuff immediately. Prolonged high pressure (cuff pressure >300mmHg or
constant pressure >15mmHg for more than 3 minutes) applied to the arm may lead to
an ecchymosis.
• Too frequent and consecutive measurements could cause disturbances in blood circulation
and injuries.
• Do not wrap the cuff on the same arm which other monitoring ME EQUIPMENT is
applied simultaneously, because this could cause temporary loss of function of those
simultaneously-used monitoring ME EQUIPMENT.
• Do not kink the connection tube during use, otherwise, the cuff pressure may continuously
increase which can prevent blood flow and result in harmful injury to the patient.
• The equipment is not AP/APG equipment and not suitable for use in the presence of a
flammable anesthetic mixture with air of with oxygen or nitrous oxide.
• This unit is not suitable for continuous monitoring during medical emergencies or
operations.
• This device cannot be used with HF surgical equipment at the same time.
• This device is not intended for patient transports outside a healthcare facility.
• To avoid measurement errors, please avoid the condition of strong electromagnetic field
radiated interference signal or electrical fast transient/burst signal.
• The operator shall not touch output of batteries/adapter and the patient simultaneously.
• Manufacturer will make available on request circuit diagrams, component parts list etc.