All manuals and user guides at all-guides.com
Monitor Components
Measurement Principle
This product uses the Oscillometric Measuring method to detect blood pressure. Before
every measurement, the unit establishes a "zero pressure" equivalent to the air pressure.
Then it starts inflating the arm cuff. Meanwhile, the unit detects pressure oscillations
generated by beat-to-beat pulsatile, which is used to determine the systolic and diastolic
pressure, and pulse rate.
The device also compares the longest and the shortest time intervals of detected pulse
waves to mean time interval then calculates standard deviation. The device will display a
warning signal with the reading to indicate the detection of irregular heartbeat when the
difference of the time intervals is over 25%.
Authorized Component
Please use the MeasuPro authorized adapter. (Not included)
Adapter Type: KH0601000UW
Input: 100~240V~50/60Hz, 0.4A Max
Output: 6V
1A
FCC Statement
This device complies with Part 15 of the FCC Rules. Operation is subject to the following
two conditions: (1) this device may not cause harmful interference, and (2) this device must
accept any interference received, including interference that may cause undesired operation.
5