16.2 Specifications
All values stated are nominal.
Height: 16.5 cm overall
Depth: 18.6 cm
Width: 32 cm
Weight: 3.57 kg
Electrical Supply: 120V 0.3A 60Hz
Fuses: T1AH 250V x 20 mm (Antisurge)
Standards: EN/IEC 60601-1, EN/IEC 60601-1-2, UL 60601-1,
CAN/CSA C22.2 No. 601.1-M90, FCC Part 15
Class 1 Type B device
Output Impulse Pressure: 60 - 200 mmHg
Impulse Duration: 1 second / 3 second (3 second factory set default)
Impulse Interval: 20 - 50 seconds (20 seconds factory set default)
Operating conditions: Temperature 0°C - +45°C
Transportation and
storage conditions: Temperature -40°C to +70°C
Relative humidity: 10%-95% condensing
Atmospheric pressure: 500mbar-1060mbar
16.3 Impulse Pressure Conversion Table mmHg/KPa
mmHG
KPa
60
8.0
70
9.3
80
10.7
90
12.0
100
13.3
17. Guidance and Manufacturer's Declaration - Electromagnetic Emission
Warning: Medical electrical equipment needs special precautions regarding EMC and needs to be
installed according to the EMC information provided. Careful consideration of this information is
essential when stacking or collocating equipment and when routing cables and accessories.
Warning: RF mobile communications equipment can effect medical electrical equipment.
EN-14
mmHG
KPa
mmHG
KPa
110
14.7
160
21.3
120
16.0
170
22.7
130
17.3
180
24.0
140
18.7
190
25.3
150
20.0
200
26.7
A-V Impulse System
Guidance and manufacturer's declaration - electromagnetic emissions
The A-V Impulse 6000 Series Controller is intended for use in the electromagnetic environment specified below.
The customer or the user of the A-V Impulse 6000 Series Controller should assure that it is used in such an environment.
Emissions test
Compliance
RF emissions
Group 1
CISPR 11
RF emissions
Class B
CISPR 11
Harmonic emissions
Class A
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
Complies
IEC 61000-3-3
Guidance and manufacturer's declaration – electromagnetic immunity
The A-V Impulse 6000 Series Controller is intended for use in the electromagnet environment specified below. The customer or the
user of the A-V Impulse 6000 Series Controller should assure that it is used in such an environment.
Immunity test
IEC 60601
Test level
Electrostatic discharge
±6 kV contact
(ESD) IEC 61000-4-2
±8 kV air
Electrical fast transient/
±2 kV for power supply lines
±1 kV for input/output lines
burst IEC 61000-4-4
Surge
±1 kV differential mode
±2 kV common mode
IEC 61000-4-5
Voltage dips, short
<5 % U
interruptions and voltage
for 0.5 cycle
variations on power supply
40 % U
input lines IEC 61000-4-11
for 5 cycles
70 % U
for 25 cycles
<5 % U
for 5 sec
Power frequency (50/60 Hz)
3 A/m
magnetic field
IEC 61000-4-8
NOTE: U
is the a.c. mains voltage prior to application of the test level.
T
A-V Impulse System
Electromagnetic environment – guidance
The A-V Impulse 6000 Series Controller uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to cause any interference in
nearby electronic equipment.
The A-V Impulse 6000 Series Controller is suitable for use in all establishments,
including domestic establishments and those directly connected to the public low-
voltage power supply network that supplies buildings used for domestic purposes.
Compliance level
Electromagnetic environment - guidance
±6 kV contact
Floors should be wood, concrete or ceramic tile.
If floors are covered with synthetic material, the
±8 kV air
relative humidity should be at least 30 %.
±2 kV for power supply lines
Mains power quality should be that of a typical
±1 kV for input/output lines
commercial or hospital environment.
±1 kV differential mode
Mains power quality should be that of a typical
±2 kV common mode
commercial or hospital environment.
(>95 % dip in U
)
<5 % UT (>95 % dip in UT)
Mains power quality should be that of a typical
T
T
for 0.5 cycle
commercial or hospital environment. If the
(60 % dip in U
)
40 % UT (60 % dip in UT)
user of the A-V Impulse 6000 Series Controller
T
T
for 5 cycles
requires continued operation during power
(30 % dip in U
)
70 % UT (30 % dip in UT)
mains interruptions, it is recommended that
T
T
for 25 cycles
the A-V Impulse 6000 Series Controller be
(>95 % dip in U
)
<5 % UT (>95 % dip in UT)
powered from an uninterruptible power supply
T
T
for 5 sec
or a battery.
3 A/m
Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or hospital environment.
EN-15