CLEANING AND DISINFECTING

WARNINGS:
- Cleaning and disinfecting are separate processes. Cleaning must
precede disinfection. Before use by a different individual: clean,
disinfect, and check the product for damage and confirm that all
package contents are provided with the product.
- Clean the product regularly and when it becomes soiled.
- Follow all manufacturer bleach container and germicidal disinfectant
safety guidelines. For a list of acceptable disinfectants, contact the
equipment provider or local health authority.
- After cleaning and disinfecting, and before using, check all components
for damage, and replace if necessary.
Cover Care Instructions
CAUTION: Frequent heat drying and bleaching will cause fabric to
break down. Air drying and cleaning without bleach are recommended
whenever possible to help extend the life of the cover.
To Identify the covering:
Refer to the cover care label to find the model number:
STRETCH-AIR Cover includes S in the model
number. Example:
Incontinence Liner model number begins with BINC.
Surface Cleaning and Disinfecting for COMFORT-TEK Cover
and Incontinence Liner Only: Spray with common household or
commercial antibacterial cleansers or disinfectants. Wipe with
a damp, water-only cloth after disinfecting. Let dry completely
before use.
Cleaning and Disinfecting (All Fabrics)
CAUTION: Remove the cover from the foam and any
inserts before cleaning or disinfecting.
To clean the cover: When possible, hand wash in warm
water with mild detergent, rinse thoroughly with clean water,
and air dry (let dry completely before use). Or machine wash
warm, using common laundry detergent and tumble dry low.
DO NOT steam clean or autoclave. DO NOT iron.
To disinfect the cover: Machine wash in hot water, using
common laundry detergent. Or hand wash warm, using 1 part
household liquid bleach per 8 parts water. Keep the product
wet with the bleach solution for 10 minutes, then rinse
thoroughly. Tumble dry low or air dry.
FOAM CARE INSTRUCTIONS on the next page
"M2-SA-18W/18D."
COMFORT-TEK Cover includes F in the model
number: Example:
.
8
"M2-FA-18W/18D."
40C
or
40C
60C
or
+
40C