To remove the rotor, use the thumbs of both hands to squeeze the "B" bracket and
then pull the rotor upwards.
Caution: The use of excessive force is forbidden when removing the rotor!
3.3.

1. Working with the unit

The instrument's cover will open when the button (3) is pressed gently.
2. Prepare test tubes with the preparation for centrifugation and place them in
the unit so that they meet the following requirements:
- the tube should be filled to a maximum of 70% of its total volume,
- The sample tubes facing each other must be of equal weight.
- place the tubes symmetrically in the rotor. An example of tube placement is
shown in the figure below.
For a round rotor:
For a rectangular rotor:
- Always verify the condition of the tubes being used. Do not place cracked or
deformed tubes in the unit!
Working with the unit