● It is only possible to connect one of the two
handpieces at the same time, so you must select
the one you need according to the treatment to be
performed.
POWER AND CONFIGURATION
● Switch on the button
the machine. The power indicator
blue. The machine will be in standby mode and
ready for operation. On the control panel:
- The timer will show 15 minutes.
- The power indicator will show 1.
- The normal skin indicator N will light up green
- The sensitive skin
indicators will light up orange.
- The cation conductor indicator
green.
- The anion conduction indicator
orange.
- The Start/Stop indicator
● Adjust the timer using the buttons
the operating time of the machine, in a range
between 1 and 30 minutes.
● Adjust the power using the buttons
the power of the machine, in a range between 1
and 9.
● Skin type selection: When the N indicator lights up
green, the selected skin type is "normal".
- When the
indicator lights up green, the
selected skin type is "sensitive".
- When the
indicator lights up green, the
selected skin type is "very sensitive".
- When a skin type is selected and lights up green,
the indicators of the other skin types light up
orange.
● Ion conduction selection:
- When the
indicator lights up green, it
indicates anion conduction (negative charge).
- When the
indicator lights up green, it
indicates cation conduction (positive charge).
- When one polarity is selected and lights up green,
the other polarity lights up orange.
● Use the
button to start or stop the device.
- When in operation, the indicator
green.
- When paused or on standby, the indicator
light up orange.
OPERATION
● Connect the device and leave it in standby mode
with the indicator
4
located on the front of
will light up
and very sensitive skin
will light up
will light up
will light up orange.
or
to set
or
to set
will light up
lit up orange.
● Apply conductive gel or cream after cleaning the
skin to be treated.
● Select the type of skin to be treated, the operating
time and set the power to minimum. Select the
polarisation of the probe using
● The patient must hold the metal bar with his/her
hand.
● Hold the hand-piece of the galvanic probe with your
hand and press
light up green.
● Start to slowly move the probe and gradually adjust
power until reaching the treatment level, always
maintaining a slight stinging sensation.
Bear in mind:
● When the polarity is set for positive electrodes,
there will be a stinging sensation given off by the
roller probe that is stronger than the one given off
by the negative electrode. Select it according to
your needs.
● The selection of ion conduction to be used should
coincide with the pH of the product applied.
- When the probe acts as a positive charge, the
conductive solution must be acidic (+).
- When the probe acts as a negative charge, the
conductive solution must be basic (-).
- If the pH is not indicated in the solution to be
used, use a pH indicator strip to determine it.
● If the Start/Stop indicator flashes, it means that the
circuit is not closed and there is no ion conduction.
Verify that the patient is holding the metal bar.
CLEANING AND MAINTENANCE
● Before cleaning the device, unplug it from the
mains.
● Clean the device with a soft, dry cloth. Do not use
liquids to clean it.
● After each use, the various components and
accessories must be sterilised and left to dry.
● For maintenance or repairs of any part of the
product, please contact your distributor or
specialised technician.