Assembly

Filling the cap
S Open the electrolyte bottle and
with electrolyte
unscrew the nozzle.
S Squeeze out excess air.
S Fill the membrane cap with electrolyte
avoiding air bubbles:
Place the electrolyte bottle
completely over the membrane cap
(see figure 2) and slowly squeeze
the electrolyte out of the bottle
in one movement.
At the same time, pull the bottle
back steadily.
The cap is completely full when
the electrolyte can be seen at
the lower thread.
Assembling
S Place the electrode adaptor on the full membrane cap in a vertical
the sensor
position.
S Do not touch the rubber seal with your fingers.
S Turn the membrane cap by hand as far as it will go.
Excess electrolyte will escape through a hole below the rubber seal
in the groove of the membrane cap while you are turning it.
S Wipe away any electrolyte with a soft paper towel or other similar item.
S Thoroughly remove gel from the grommet using clean, warm water.
ProMinent
®
CAUTION
Wear safety goggles and protective clothing when handling
water or solutions that contain chlorine.
TAKE CARE
• Do not touch or damage the membrane or electrodes.
• The electrolyte is susceptible to oxidation: always keep
the electrolyte bottle sealed after use. Do not put the
electrolyte into other transparent containers.
• The electrolyte should not be stored for more than 1 year
and should not be yellow (see the label for the date of
manufacture).
• Avoid air bubbles when pouring the electrolyte.
• The membrane cap must be used only once.
Assembly
Figure 2: Membran cap