5.2.8
Setting the pulse per liter
For service and expansion purposes only!
en
Only change the setting if instructed to do so by Renfert!
Key
6

Cleaning / Maintenance

Safe functioning of the system is only possible if the filter cartridge is replaced regularly. The replacement
cycles depend on the carbonate hardness of the drinking water, usage, and the blend level set�
We recommend replacing the filter cartridge after 6 months, but no later than after 12 months.
The operator is required to check the system for leaks on a daily basis�
When replacing the filter cartridge, all parts must be checked for contamination and damage. Damaged
parts must be replaced and contamination removed�
If the system is not used for an extended period, the filter cartridge must be replaced.
6.1
Flushing the system
The system must be flushed thoroughly after maintenance work or if it has not been used for an extended
period�
⇒ Run the flushing hose into a suitable container (e.g., a bucket) or into a drain.
⇒ Open the purge valve (5, Fig. 5).
⇒ Turn on the water supply�
⇒ Allow the quantity of water specified in the table below to flow through.
⇒ Close the purge valve (CLOSE).
Filter cartridge M for POWER
steamer water softener
Filter cartridge L for POWER
steamer water softener
⇒ Then carry out 3 rinsing cycles using the POWER steamer�
6.2
Spare parts
You can find wear parts and spare parts in the spare parts list available online at
www�renfert�com/p918�
Enter the following item number: 18460100�
The components excluded from the warranty (wear parts, consumables) are
marked in the spare parts list�
Press the "MENU" key until the settings are shown�
- Display: Settings�
Press and hold the "MENU" key for approx� 5 seconds�
- "L" or "Gal" flashes.
Press and hold the "MODIFY" key for approx� 10 seconds�
- Pulse per liter is shown, the active digit flashes.
Press the "MODIFY" key:
- Set the active digit�
Press the "MENU" key:
- Confirm the active digit and move to the next digit.
- The active digit flashes (repeat the procedure until the last digit).
Flush volume after 1 week of
stagnation:
3 liters
5 liters
- 12 -
Display
Flush volume after 4 weeks of
stagnation:
15 liters
25 liters