Graph of test tube heating after 30 minutes of centrifugation
ROTOR 6M .01 / 6M .06
ROTOR 6M / 6M .05
ROTOR 6M .02

Error code table

Error code
Cause
No main power connection .
Power failure
Centrifuge lid is left open .
Total imbalance of test tubes is more than
7 gram .
Processor Control Board is overheated .
Residual rotation of the rotor .
Centrifuge CM-7S Plus
2500
3000
3500
2
3
3
1 .5
2
2 .5
2
3
3
USER MANUAL
6М.02
6М / 6M.05
6М.01 / 6M.06
3 .5
6
7
8
3
5
6
7
4
6
8
9
Solution
Check connection of the power cable .
Close and press on the lid till you hear a
click .
Load rotor symmetrically .
Give a little time and centrifuge will cool
down automatically .
Wait until the rotor is motionless .