Cleaning / Lubrication / Checking of the Attachment, Slim Attachment Hub and Slim Tube
Manual Cleaning without Cleaning Tools
_ Pre-cleaning
Remove contaminants on the surface with a cloth or soft bristle brush under running tap water.
For products with moving parts, perform cleaning while moving all parts (see "2 Moving Parts of Each Product").
≤ 38°C, ≥ 3.5 L/min, ≥ 30 sec.
Water quality: Drinking water
_ Cleaning
1
Prepare cleaning solution in the cleaning tank under the
following conditions (prepare the cleaning solution according to
the instructions of the detergent manufacturer).
Temperature
Water quality
Distilled water
Deionized water
45°C
Demineralized
water
* Detergent used for validation:
neodisher MediZym (DR. WEIGERT)
2
Immerse the product in the cleaning tank for 20 - 30 min.
Detergent
Neutral enzymatic
detergent for medical
purposes (pH 7.0 - 8.0)
13
3
Thoroughly rinse the dirt released in immersion as well as the
detergent under running tap water.
For products with moving parts, perform cleaning while moving
all parts (see "2 Moving Parts of Each Product").
≤ 38°C, ≥ 3.5 L/min, ≥ 10 sec.
Water quality: Drinking water
4
Rinse the product surface with distilled, deionized or
demineralized water.