Descargar Imprimir esta página

Cleaning / Checking The Irrigation Nozzle; Cleaning - NSK Primado2 Manual Del Usuario

Ocultar thumbs Ver también para Primado2:

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

Cleaning / Checking the Irrigation Nozzle

9
9 – 1 Cleaning
Check the compatible Cleaning Adaptor before using the Cleaning Tools.
Product
Order Code
P200-IN-1A155, P200-IN-1A175
PD-IN-D, PD-IN-2ES, PD-IN-2S, PD-IN-2M
Irrigation Nozzle
PD-IN-2L
PD-IN-3S,PD-IN-3M,PD-IN-3L
P200-IN-BS
9 – 1 – 1 Manual Cleaning
Manual Cleaning with Cleaning Tools
Use the Cleaning Tools according to the following system figure.
Irrigation Nozzle
_ Pre-cleaning
Remove contaminants on the surface with a cloth or soft bristle brush under running tap water.
≤ 38°C, ≥ 3.5 L/min, ≥ 15 sec.
Water quality: Drinking water

_ Cleaning

1
Mount the Cleaning Adaptor IN and the Irrigation Nozzle to the
Cleaning Tube 1.5/L25.
Pull the Cleaning Adaptor IN to make sure that it is securely
connected.
* If the Cleaning Tube 1.5/L25 is deteriorated, replace it with a
new one.
Cleaning Tube 1.5/L25
Cleaning
Cleaning Adaptor IN
2
Screw the disposable luer-lock syringe (hereafter referred to as
syringe) into the Cleaning Adaptor IN.
23
Compatible Cleaning Adaptor
Cleaning Adaptor IN
PD-CTAD-IN
Drying
Luer-lock syringe
Air Gun

Publicidad