FAULT INDICATOR
If a fault should occur with your wheelchair when switching it on or during a journey, the protection system will
automatically activate and send fault signals by means of a flashing light on the hand control. At this point, you should
not use your wheelchair until the problem has been resolved.
Number of
blinks
1
2
3
4
5
6
7
8
9
10
PRECAUTIONS
1. The electric wheelchair is suitable for indoor and outdoor use.
2. Do not use on bumpy roads or climb steps frequently.
3. It is not suitable to climb long slopes for long periods of time as this can cause the engine to overload and the battery
to drain quickly.
4. If you need to stop the wheelchair, reset the control lever and it will stop automatically.
5. Do not disassemble the battery under any circumstances.
6. If you need to fold the wheelchair, the battery must be disconnected and removed.
7. Caution. Flammable material, avoid exposure to heat sources such as flames or sparks. Avoid transporting batteries
close to flammable or combustible objects.
8. Should you need to brake in an emergency situation, pull the steering lever in reverse and the wheelchair will stop
instantly thanks to the system's electromagnetic brake.
9. Keep tools and other metal objects away from the edges of the battery, as short circuits or electric shocks may occur
in case of contact and cause injury.
11. Extreme caution must be exercised when driving in reverse. The drive wheels could hit an obstacle and cause the
wheelchair to lose control and tip over.
Fault
Faulty contact to charge the battery or
batteries too low.
Faulty contact of the left motor circuit.
Shortcircuit in the left motor.
Bad contact of the right engine circuit.
Shortcircuit in the right motor.
The controller has stopped working for
safety reasons.
The steering lever of the control is faulty.
The control has an internal fault.
The right and left brakes or their circuit
is faulty.
Battery overcharge.
ES
EN
FR
IT
Check the cables to verify that the battery
is properly charged. Charge the battery.
Check the connection and wiring between
the control unit and the left motor.
The left side motor is short-circuited,
consult a technician.
Check the connection and wiring between
the control unit and the right-hand motor.
The right side motor is short-circuited,
consult a technician.
An external signal has interfered with the
controller, check if the charger is connected.
Do not use the controller while the battery
is charging.
Move the lever to the neutral position and
reset the control.
Consult with a technician.
There is a short circuit or damage to the
electromagnetic brake circuits. Consult a
technician.
Check the battery, its connections and
wiring.
DE
PT NL
SWE
Solution
PL
DK