3. Then rinse the parts in clean water.
4. Vigorously shake the lower part to remove any water that may have entered the heater
rod contact. Dry the parts thoroughly with a soft cloth. The heater rod contact must be
dry.
5. Check whether the pressure measurement tube is still correctly fitted in the upper part.
If necessary reinsert it.
For hygiene reasons, we recommend that the plastic parts are replaced after a maximum
of two years use.
CLICK 2 drying
If water penetrates into the pressure line, then dry CLICK 2 as follows:
Caution!
The bellows must only be used for drying the humidifier. During drying CLICK 2 must
not be connected to the therapy unit. Otherwise the unit could be damaged.
2. Press the bellows together so that any moisture is blown out.
28
Hygienic preparation
To seal the pressure line at the inlet connections, push
the thin end of the closure by a slight rotational move-
ment into the seal of the pressure line.
Clean the upper and lower parts with lukewarm water to
which a little detergent has been added. We recom-
mend cleaning the housing parts once a week in the
upper tray of the dishwasher at a maximum tempera-
ture of 65 °C.
6. Fit the upper and lower parts together such that the
marks are aligned. Turn the upper and lower parts
against each other clockwise as far as they will go.
7. Remove the closure. The humidifier is now ready for use
again.
Marks
1. Push the bellows included in the delivery, as illustrated,
onto the end of the pressure line at the outlet connec-
tors.