INSTRUCTIONS FOR USE OF REMOTE CONTROL
The remote control is connected to the control unit by a spiral extendable cable
Keys:
1. UP" key for lifting
2. DOWN" key for lowering
Visual indications:
3. LED for max. load exceeded: the lift moviment is interrupded if the
4. Led for monitoring and signaling the battery charge status

13. WARNINGS

• The lift is a device designed for moving patients with motor problems
for short distances, for instance from a bed to a wheelchair or comfortable chair,
and back. It must not be used for transporting the patient.
• While lifting the patient, make sure the base legs are as far apart as possible,
the control lever is securely locked and the rear wheel brakes are secured.
• While moving the patient, please avoid crossing barriers that might cause
the lift to tip over.
• The use of the patient lifts described in this manual, in combination with the Moretti harnesses
provided among the accessories, it allows to pick up and lift patients off the ground.
WARNING!
Always check that the cradle is in good condition. Do not use cradles showing
signs of damage or wear.

14. CLEANING

To clean the cradle and the structure use a solution of alcohol and neutral soap (in equal parts)
mixed with water.

15. MAINTEINANCE

It is extremely important to check the lift before each use.
More specifically, please check that all screws and bolts are well tightened.
Please, check the lift can move freely with the wheels unlocked by the brakes.
Check the rear wheel brakes function correctly.
It is very important to regularly check the connection pins and stop rings are
correctly located (the pins connect the actuator to the upper and lower part of the
lift, and the lifting beam where you hook the cradle).
The mechanical parts must be lubricated with a drop of oil in order to facilitate their
sliding and resistance to wear. In particular, pay attention to the mechanical retractor
for the legs of the patient lift. Lubricate as indicated above every two or three months.
load exceeds the max. load treshold and the control unit generates an
acoustic signal.