Cleaning the micro-motor (optional accessory)
Remove all attachments to the micro-motor prior to cleaning. The outer surface of the micro-
motor should be cleaned with an antiseptic solution. Never oil the micro-motor. Ensure that
no moisture gets inside the micro-motor.
WARNING: Do not dip the micro-motor and cord directly into sterilization fluids.
chemicals will interfere with normal operations if it reaches the inside of the system.
WARNING: Do not leave disinfectants on parts longer than recommended by the
manufacturer of the disinfectant to avoid damaging materials.
NOTE: The micro-motor should be cleaned after each use between patients.
Cleaning the straight hand piece (straight nose cone) attachment (optional accessory)
Clean the straight hand piece preferably in an ultrasonic tank and rinse in distilled water.
Dry the straight hand piece.
Disinfect the straight hand piece with a cloth and mild disinfectant containing a neutral pH.
Seal the straight hand piece inside a sterilization pouch. If you are autoclaving more than 1
tip, place each straight hand piece in an individual bag. You may also use the tip holder that
comes with the KRUUSE-SP2 in place of the sterilization pouch.
Autoclave the straight hand piece .
Cleaning the Prophy Head (unless disposable) (optional accessory)
The prophy head should be cleaned to remove hair and debris approximately after every 10 dental
polishing.
1. Remove prophy angle from handpiece and rubber cup.
2. Remove head gears by turning knurled knob clockwise ( threads are reversed) turn to your right.
3. Clean with fine brush and lubricate head gears with mineral oil, Vaseline or any non-toxic
lubricant.
4. Place head gears back and tighten by turning Left or counterclockwise.
5. Normal polishing speeds are around 2,000 to 3,000 RPM, exceeding this may damage your
prophy angle.