If the installation of your ART-SP6 system is performed by non-Bonart dealer personnel, check to
see that the requirements below are followed.
Water Line Requirements
 Incoming water supply line pressure to the Scaler must be 25 psi (172 kPa minimum) to 60
psi (414Kpa maximum). If your dental water supply line pressure is above 60 psi, install a
water regulator on the water supply line to your ART-SP6 unit.
 After installing the water supply system, flush the water thoroughly prior to connecting to
the scaler.
 Manually shut off the water valve of your dental water supply system pior to leaving the
office unoccupied and unattended.
Electrical Requirements
Refer to Section X: Specifications