Safety notes
Control unit U8300
> Never fill or top up the coolant tank with liquids hotter than 30°C/86°F.
> Replace a faulty or leaky pump cartridge immediately.
Control unit U8200, U8300
Risks due to electromagnetic fields
The functionality of implantable systems, such as cardiac pacemakers and implantable cardioverter defibrillators (ICD)
can be affected by electric, magnetic and electromagnetic fields. This device complies with the reference values
defined in EN 50527-2-1/2016 for unipolar and bipolar pacemakers and is therefore suitable for patients with
pacemakers.
> Find out if patient and user have implanted systems before using the device and consider the application.
> Keep a safe distance of at least 10 cm (3.94 inch) between the device and the cardiac pacemakers.
> Make appropriate emergency precautions and take immediate action on any signs of ill-health.
> Symptoms such as raised heartbeat, irregular pulse and dizziness can be signs of a problem with a cardiac pacemaker
or ICD.
The control unit is designed for use with the iM3 handpiece U8205 so only this is to be used with the control unit. The use
of other handpieces could lead to a malfunction or destruction of the electronics.