Symptoms:
4.
Forefoot feels too rigid
Forefoot feels too long
Feels like walking up hill
Vacuum System
Symptoms:
1.
Unable to generate a
vacuum
2.
Unable to maintain a
vacuum
All manuals and user guides at all-guides.com
1. Reduce dorsiflexion resistance
2. Check A-P shift alignment;
ensure foot is not too anteriorly positioned
3. Check distribution of plantar flexion and
dorsiflexion movement; ensure that there is
sufficient dorsiflexion range
4. Check the spring category is not too rigid for the
weight and activity of the user, if so fit a lower
rate spring
Vacuum tube(s) split or disconnected
Inspect and repair/replace as necessary
Check and clean/replace the check valve
Filter blocked, replace Filter
Limited ankle movement creating insufficient vacuum due to:
1. Excessive PF/DF setting
2. Footwear
Vacuum tube(s) split or disconnected
Inspect and repair/replace as necessary
Check and clean/replace the check valve
Leakage at socket valve/barbs
Reseal valve/barb
Porous socket
Seal with lacquer/re-make
Check the integrity of the vacuum seal at the socket/
residuum interface
13
Remedy
Cause/Remedy
938367/1-0816