Low Level Disinfection (optional)
1. Disinfect the device in a disinfecting solution that has one of the following
active Ingredients:
≥ 2.4% glutaraldehyde
of 45 minutes at 25 °C
or
≥ 3.4% glutaraldehyde
of 20 minutes at 25 °C
or
≥ 0.55% ortho-phthalaldehyde
of 12 minutes at 25 °C.
2. Prepare the disinfecting solution according to the manufacturer's
instructions.
3. Per manufacturer's recommendations, immerse the device, filling all
lumens, in the disinfecting solution for the required time at the appropriate
temperature.
4. Thoroughly rinse and flush all parts and lumens with running,
demineralized water to remove the disinfectant.
5. Dry all parts with a lint-free towel immediately after rinsing.
CIDEX Activated® is validated for disinfection efficacy.
4
CIDEX Plus® is validated for disinfection efficacy.
5
CIDEX® OPA is validated for disinfection efficacy.
6
Drying
For automated drying, use the drying cycle provided with the washer/
disinfector.
For manual drying, use a lint-free cloth.
Dry any lumens with compressed air.
Maintenance, Inspection, and Testing
Inspect the device on a continual basis. If a problem is observed
or suspected, the device should be returned for repair.
Inspect all components for cleanliness. If fluid or tissue buildup
is present, repeat the above cleaning and disinfection procedures.
Inspect the camera cable for cuts and breaks. Return any damaged
urology camera to Stryker for service.
Packaging
N/A
18