Insufficient blood
sample.
Defective test strip.
The test does
not start after
Sample applied
applying the
after the device
sample.
is automatically
turned off.
Defective device.
Error in performing
the test.
Control solution
vial was poorly
shaken.
Expired or
The control
contaminated
solution
control solution.
testing result
is out of
Control solution
range.
that is too warm or
too cold.
Defective test strip.
Device
malfunction.
Repeat the test using a
new test strip with larger
volume of blood sample.
Repeat the test with a new
test strip.
Repeat the test with a
new test strip. Apply
sample only when
flashing " " appears on
the display.
Please contact customer
service.
Read instructions
thoroughly and repeat
the test again.
Shake the control
solution vigorously and
repeat the test again.
Check the expiration
date of the control
solution.
Control solution, device,
and test strips should
be at room temperature
(20°C to 25°C / 68°F to
77°F) before testing.
Repeat the test with a
new test strip.
Please contact customer
service.
EN-22
Specifications
1000 measurement results with
Memory
respective date and time
Dimensions
89.8 (L) x 54.9 (W) x 18 (H) mm
Power Source
One 1.5V AAA alkaline battery
Weight
46.1 g (without battery)
External output
Bluetooth
Auto electrode insertion detection
Auto sample loading detection
Auto reaction time count-down
Features
Auto switch-off after 3 minutes without
action
Temperature warning
8°C to 45°C (46.4°F to 113°F), below 85%
Operating Condition
R.H. (non-condensing)
Storage/
-20°C to 60°C (-4°F to 140°F), below
Transportation
95% R.H
Condition
Blood Glucose Test: mg/dL or mmol/L
Hematocrit Test: %
Hemoglobin Test: g/dL
Measurement Units
β-ketone Test: mmol/L
Total cholesterol Test: mg/dL
Uric acid Test: mg/dL