Descargar Imprimir esta página

Vermeiren LiNX 200 Manual De Instrucciones página 8

Ocultar thumbs Ver también para LiNX 200:

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 63
EN
NL
4 Operating console functions
The operating console enables you to control all driving components as well as the motors to
adjust your body position. For your own safety, and to prevent that the wheelchair gets damaged,
movements are monitored by the electronics. This may result in block of movement, speed
decrease or complete stop.
4.1 Switching ON / OFF
Put the joystick in the neutral (central) position before pressing the ON/OFF button.
If the console is switched on with the joystick out of the neutral position, the wheelchair will
not drive. Release the joystick (return to neutral) within 5 seconds to unlock the drive. If the
joystick is not released within 5 seconds, the fault turns into a Module fault which is
indicated by the red blinking LED in the ON/OFF button. The wheelchair will not drive. To
clear the Module fault, release the joystick and switch the operating console OFF and then
ON again.
To switch the operating console ON, press the ON/OFF button once.
The LED (status indicator) in the ON/OFF button goes on. When it turns green, the operating
console is ready for operation. When the LED starts blinking in red, there is a failure, see
§4.2 and §5.
To switch the operating console OFF when it is ON: press the button once.
The operating console will power down and the status indicator goes out.
4.1.1 Emergency stop
If you need to stop a driving or body-positioning movement quickly, press the ON/OFF button to
perform an EMERGENCY STOP. The wheelchair will come to a halt quickly.
4.1.2 Sleep mode
The operating console will go into sleep mode after a period of inactivity (default 5 minutes).
When the system sleeps, it is partially turned off to reduce energy consumption.
To reactivate the operating console, press a button or move the joystick.
4.2 Failures and blocks
When the operating console is powered and a failure appears, the ON/OFF button
starts blinking in red. The number of blinks indicates the fault code. In case of a failure,
one or all functions of the operating console are blocked. For more information, see
§5.
4.3 Battery status and charging
WARNING
Follow the charging instructions in the wheelchair instruction manual and the suppliers' manual
of the battery charger.
Risk of injuries and damage
6
LiNX 200
2023-03
Operating console functions

Publicidad

loading