Descargar Imprimir esta página

Xylem SI Analytics Lab 745 Manual De Instrucciones página 53

Equipo de mesa para oxígeno, ph, redox, ise y conductividad
Ocultar thumbs Ver también para SI Analytics Lab 745:

Publicidad

Idiomas disponibles

Idiomas disponibles

6
Info
6.1
General
In this menu item, the first page shows the information about the current calibration values and the last
calibration time (date).
The second page shows the maximum and minimum values of the main measured value and the time
when these values are measured.
On this page, you can press the (CONF) button to reset the minimum and maximum values.
6.2
Menu structure info
CAL
INFORMATION
7
Data communication / printer
PC-Software „Labx45 pilot"
7.1
The reading of the data logger as well as the configuration of the measuring system can also take
place via the software "Labx45pilot".
Use only interface cable Z 613 to transfer data from the meter to the PC. This cable is not included
in the delivery and must be purchased separately ( chapter 12 Accessories).
You will find the software here:
http://www.si-analytics.com/downloads/software-updates.html
To use this function, press (CONF). This will take you to the menu item "Configuration" and on to the
submenus item "protocol" choose "standard".
7.2
Serial data output via ASCII
The current measured values can be cyclically output via the serial interface as ASCII characters.
Use only interface cable (Interface HMG USB) to transfer data from the meter to the PC. This cable
is not included in the delivery and must be purchased separately ( chapter 12 Accessories).
To use this function, press (CONF). This will take you to the menu item "Configuration" and on to the
submenus item "protocol" choose "ASCII" and setup the "cycle time".
The interface setting of the receiver is 4800 bauds, 1 stop bit and no parity.
All output characters are in the ASCII format.
Data decription: dd mm yy hh minmin E1E1 MMMM E2E2 NNNN E3E3 TTTT
Dd
2 Byte
Day
mm
2 Byte
Month
yy
2 Byte
Year
hh
2 Byte
Hour
minmin
2 Byte
Minute
E1E1
4 Byte
Unit
MMMM
4 Byte
measured main value
E2E2
4 Byte
Unit
NNNN
4 Byte
Secondary measured value
E3E3
4 Byte
Unit
TTTT
4 Byte
temperature
MIN/MAX INFO
53
MESSMODUS
RESET MIN/MAX

Publicidad

loading

Este manual también es adecuado para:

Si analytics lab 845Si analytics lab 945