Bare Fiber Universal, Purged, Single Use
Manually screw on SMA-X connector fully and tightly onto the laser fiber socket of the laser
surgery unit (do not use any tools).
The red laser status indicator and the information line Connect fibre on the surgical laser
unit disappear.
The purged laser fiber can also be used without AIR (purge air).
WARNING
Danger of tissue damage due to compressed air (AIR)!
Ensure that the air pressure of the AIR (purge air) on the Luer Lock connection for the
laser fiber does not exceed the maximum value of 5 bar.
Ensure that the quality of the purge air meets the requirements of the intended surgery.
Connect the AIR purge air hose on the proximal laser fiber end to the purge air supply unit
on the laser surgery unit via an AIR sterile air filter.
Check the AIR purge air stream. A slight air flow must be felt at the distal end of the laser
fiber.
Check the purged laser fiber for damage on its entire length.
Activate the pilot laser and check the beam quality. The pilot laser beam should appear
round and even on a bright surface at a distance of a few centimeters.
5
Disposal
The purged laser fiber must be disposed of as infectious material after use according to the
national regulations and disposal guidelines.
6
Technical Data
Designation
REF
Fiber core diameter
External diameter
Length
Max. laser output
Purge air connection
CE conformity marking
Revision 1
Bare Fiber Universal, purged, single use, sterile
79-700-45-04
400 µm
2.1 mm
3 m
without contact 40 W
Medical compressed air (AIR), max. 5 bar