P3.2: Normalization temperature: (nor)
The meter will normalize its cond. measurement to a standard temp.
which you preset. The adjustable range is 15 to 30 °C (59 to 86 °F).
Meter default is 25 °C (77°F). When in P3.2, press ENTER again and
the default normalization temperature flashes on LCD. Press UP/DOWN
to change the value from 15.0 to 30.0 °C. Press ENTER to confirm and
enter P3.3. For more information of temperature effect on measurement,
refer to Appendix C.
P3.3: Manual temp. Compensation: (Int)
When ATC is disabled, you can manually enter the temp.value of
solution. Any temperature between 0 and 50 °C (32 to 122 °F) is
selectable. Meter default is 25 °C (77°F).
When in P3.3, press ENTER again and the default manual temperature
25.0°C blinks on LCD. Press UP/DOWN to change the value. Press
ENTER to confirm and return to P3.0.
P4.0: V iew calibration data (CAL)
Recall previous calibration data and help to know when is needed to
re-calibrate. It's for "Review" purpose only In P4.0, press ENTER repeatedly
to view P4.1to P4.5, and it returns P4.0 after 4.5 P4.1 is calibration data for
range 1, P4.2 is for range 2, .....P4.5 is for range 5. If no previous calibration
data at particular range, the display will show " - - -".
P5.0: Electrode data: (ELE)
To check the probe cell constant value for diagnostic purposes. The cell
constant is adjusted according to your calibration. In P5.0, press ENTER
repeatedly to view P5.1 to P5.5 , and it returns to P5.0 after P5.5. P5.1 is
the cell constant value for range 1. P5.2 is for range 2, .... P5.5 is for range 5.
P7.0: Reset to factory default setting (rSt)
P7.1: Meter reset (rSt)
Reset all parameters to factory default. This function will clear all
calibration data and all setup value you've done. In P7.0, press ENTER
to enter P7.1. Press UP/DOWN to select "n"-NO or "y"-YES. Press ENTER
to confirm and return to P7.0. To completely recalibrate a meter or
using a replacement probe, it is suggested to clear all calibration data
in memory.