MEASURING DC VOLTAGE (see figure 6)
Warning
To avoid harm to you or damages to the Meter from electric shock, please do not attempt
to measure voltages higher than 1000V although readings may be obtained.
The DC Voltage ranges are: 6.000V, 60.00V, 600.0V and 1000V. If over the range, it displays
"OL". To measure DC voltage, connect the Meter as follows:
1.
Insert the red test lead into the
terminal.
2.
Set the rotary switch to
3.
Connect the test leads across with the object being measured. The measured value shows
on the display.
Note:
If the value of voltage to be measured is unknown, set the Meter to autoranging or use the
maximum measurement position 1000V and reduce the range step by step until a satisfactory
reading is obtained.
In each range, the Meter has an input impedance of 10MΩ. This loading effect can cause
measurement errors in high impedance circuits. If the circuit impedance is less than or equal to
10kΩ, the error is negligible (0.1% or less).
When DC voltage measurement has been completed, disconnect the connection between
the testing leads and the circuit under test, and remove the testing leads away from the input
terminals of the Meter.
Figure 6. Measuring DC Voltage
terminal and the black test lead into the COM
31