27
FEATURES
Irregular Heartbeat Detector
The symbol
was detected during measurement.
The heartbeat rhythm that is more than or less than 25% from the average rhythm is usu-
ally defined as an irregular heartbeat rhythm.
Talking, moving, shaking or an irregular pulse during the measurement can result in the
appearance of this symbol.
Usually this is not a cause for concern, however if the symbol appears often, we recom-
mend you seek medical advice.
And please note that the device does not replace a cardiac examination, but serves to
detect pulse irregularities at an early stage.
*Note!
• The pulse display is not suitable for checking the frequency of heart pacemakers.
If a certain pulse irregularity is detected during measurement often, we recommend you
seek medical advice
• As a safeguard, we recommend that if you have arrhythmias such as atrial or ventricular
premature beats and atrial fibrillation or any other special conditions you should check
with your physician before using your device.
• The IHB function is not designed for use by people with arrhythmias nor for diagnosing
or treating an arrhythmic problem. In order to filter the unstable status of user and avoid
affecting the detection of heart rate from any movement, shaking or talking in the begin-
ning of measurement, the method of averaging heart beat intervals of subject device is
calculated with the three proper heart beat pulses detected in the beginning of measure-
ment and that is different from a strict mathematical averaging of all recorded intervals.
• At least 3 beats with at least 25% difference from the average heart beat interval will
generate the IHB icon on the screen.
will appear on screen indicating a certain heartbeat irregularity