39
Test specifications for enclosure port immunity to RF wireless communications equip-
ment.
Test frequency (MHz)
385
450
710
745
780
810
870
930
1720
1845
1970
2450
5240
5500
5785
NOTE: If necessary to achieve the IMMUNITY TEST LEVEL, the distance between the transmitting
antenna and the ME EQUIPMENT or ME SYSTEM may be reduced to 1 m, the 1 m test distance is
permitted by IEC 61000-4-3.
a). The carrier shall be modulated using a 50% duty cycle square wave signal.
b). AS an alternative to FM modulation, 50% pulse modulation at 18 Hz may be used because while it
does not represent actual modulation, it would be worst case.
Disposal: The product must not be disposed of along with other domestic waste.
The users must dispose of this equipment by bringing it to a specific recycling
point for electric and electronic equipment.
GIMA WARRANTY TERMS
The Gima 12-month standard B2B warranty applies.
Modulation
Pulse modulation 18 Hz
FM ± 5 kHz deviation 1kHz sine
Pulse modulation 217 Hz
Pulse modulation 18 Hz
Pulse modulation 217 Hz
Pulse modulation 217 Hz
Pulse modulation 217 Hz
ENGLISH
IMMUNITY TEST LEVEL (V/m)
27
28
9
28
28
28
9