TROUBLESHOOTING
SYMBOLS/SYMPTOMS
No power-on by pressing
button.
Measuring Error Symbol
appears when blood
pressure value displayed
is excessively low or high.
Measuring Error
Symbol.
Measuring Error
Symbol.
Measuring Error
Symbol.
Note: If "EP" appears on the display, just return the device to your local distributor
or importer.
ENGLISH
CONDITIONS/CAUSES
Worn-out batteries.
Battery polarities have been
positioned incorrectly.
Cuff has been placed
incorrectly.
Did you talk or move during
measurement?
Shaking of the arm with
the cuff on.
Air circuit abnormality.
Cuff tube may not be plugged into
monitor correctly.
Inflation pressure exceeding
300 mmHg.
Error determining
measurement data.
INDICATION/CORRECTION
Replace them with 4 new
AA alkaline batteries.
Re-insert the batteries in the
correct positions.
Wrap the cuff properly so that it
is positioned correctly.
Measure again.
Keep arm steady during
measurement.
Check cuff connection.
Measure again.
Switch the unit off, then measure
again.
Measure again.
34