INDICATIONS FOR USE:

Repairs Medcomp
WARNINGS:
Do NOT use to repair catheters other than those specified
above.
Do NOT replace extension set if tubing is swollen or displays
signs of degradation.
This extension set should NOT be used for any purpose other
than indicated in these instructions.
To maintain peak performance of the extension set, it is
recommended that the extension set be replaced every 6
months.
Do NOT re-sterilize the extension set or accessories by any
method.
Do NOT use extension set if package is opened or damaged.
Do NOT use extension set if any sign of product damage is
visible.
Failure to clamp could lead to air embolism or blood loss.
CAUTIONS:
In the rare event luer connector separates from any component
during use, take all necessary steps and precautions to prevent
blood loss or air embolism.
Federal Law (USA) restricts this device to sale by or on the
order of a physician.
This extension set is for Single Use Only.
The manufacturer shall not be liable for any damages caused
by reuse or re-sterilization of the extension sets.
Re-Use may lead to infection or illness/injury.
Contents sterile and non-pyrogenic in unopened, undamaged
package.
STERILIZED BY ETHYLENE OXIDE
ALL CATHETER REPAIRS ARE DONE AT THE DISCRETION OF
THE ATTENDING PHYSICIAN.
When cutting catheter to desired length, assure lumen is cut
square and that remaining catheter lumen is not damaged.
Avoid clamping near adapter or luer of the extension set.
Catheter will be damaged if clamps other than what is provided
with this kit are used.
Clamping of the tubing repeatedly in the same location may
weaken tubing.
Examine catheter extension set before and after each treatment
for damage.
To prevent accidents, assure the security of all caps and
bloodline connections prior to and between treatments.
Use only Luer Lock (threaded) Connectors with this catheter.
Repeated overtightening of bloodlines, syringes, and caps will
reduce connector life and could lead to potential connector
failure.
Split Stream
®
Catheters
®
STERILE EO
-1-