Descargar Imprimir esta página

Velp Scientifica AREX 5 Manual De Instrucciones página 9

Ocultar thumbs Ver también para AREX 5:

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 25
Note: The AREX 5 Digital always has primary control of the heating plate temperature. The temperature control
feature of the heating plate can also be used as a safety thermostat. In this case, the maximum temperature of the
heating plate will not exceed the temperature setting on the magnetic stirrer (see option
meaning that a longer heating time is required in order to reach the VTF thermoregulator temperature setting, thus
reducing temperature oscillation at a setpoint value.
Operation with external probe
Turn off the instrument using the main switch
Screw the threaded support rod into its seat on the back of the instrument (optional)
Fasten the clamp onto the support rod (optional)
Place the external temperature probe into the clamp (optional); then place it into the sample contained inside the
flask
Plug the probe into the dedicated socket on the back of the instrument. The AREX 5 Digital automatically recognize
the type of temperature probe (Pt100 or Pt1000)
Turn the unit on using the main switch. When the external probe is connected correctly, the Probe led is on
Turn the left knob to set the working temperature; the Set led indicates that the display is showing the set point
temperature
Push the left knob to start heating. The left display shows the temperature read by the probe for 3 seconds and the
setpoint for 1 second
3
chap. 4),

Publicidad

loading

Este manual también es adecuado para:

F20500580F20510580