Implementation:
- Fill a container big enough to hold all of the parts to disinfect with a solution of potable
water and disinfectant, according to the proportions indicated on the packaging of the
disinfectant.
- Completely immerse each part in the solution, taking care to avoid the formation of air
bubbles on the parts. Leave the parts immersed for the amount of time indicated on the
packaging of the disinfectant associated with the concentration chosen for the solution.
- Remove the disinfected parts and rinse abundantly with warm potable water.
- Dispose of the solution following the instructions provided by the disinfectant manufacturer.
method B: Disinfect the accessories by boiling them in water for 10 minutes; use demineralised
or distilled water to prevent calcium deposits.
method C: Disinfect the accessories with a hot steam steriliser for baby-bottle (not the
microwave type). Perform the process faithfully following the instructions of the steriliser. To
ensure that the disinfection is effective, choose a steriliser with an operating cycle of at least 6
minutes.
After having disinfected the accessories, shake them vigorously and place them on a paper
towel. Alternatively, dry them with a jet of hot air (for example, from a hair dryer).
CLEANING IN A CLINICAL OR HOSPITAL SETTING - DISINFECTION AND STERILISATION
SANITIZATION: The accessories that can be sanitized are (M0,M1,M2,R).
Before disinfection or sterilisation, sanitise the accessories, choosing one of the methods
described below.
method A: Sanitise accessories under potable hot water (approximately 40°C) with a gentle,
non abrasive dish detergent.
method B: Sanitise accessories in the dishwasher with a hot cycle.
DISINFECTION
DISINFECTION: The accessories that can be disinfected are (M0,M1,M2,R)
The disinfectant must be an electrolytic chloroxidizer (active principle: sodium hypochlorite)
specific for disinfecting, which is available in any pharmacy.
Implementation:
- Fill a container big enough to hold all of the parts to disinfect with a solution of potable
water and disinfectant, according to the proportions indicated on the packaging of the
disinfectant.
- Completely immerse each part in the solution, taking care to avoid the formation of air
bubbles on the parts. Leave the parts immersed for the amount of time indicated on the
packaging of the disinfectant associated with the concentration chosen for the solution.
- Remove the disinfected parts and rinse abundantly with warm potable water.
- Dispose of the solution following the instructions provided by the disinfectant manufacturer.
If you want to also perform the STERILISATION, jump to the STERILISATION paragraph.
After having disinfected the accessories, shake them vigorously and place them on a paper
7