Resistance Decade Boxes / English
4. Range selection
The 28 front panel switches are used to select the resistance that will be available on the
red and black terminals. When a switch is set to the IN position, the value printed above the
switch is added to the total resistance available at the posts. If all of the switches are set to
out, the total output resistance is zero (0.3Ω internal resistance-approx).
For example, if the desired output value is 35.3kΩ, set the following switches to the IN
position: 30k, 4k, 1kΩ and 300Ω.
5. Testing
This device can be used to verify the calibration integrity of MultiMeters, LCR meters,
calibrators, etc. connect as described above in the binding post connection section. set the
resistance switches to output the desired resistance. Ensure that the voltage supplied by the
device under test does not overload the 0.3W internal resistor power rating. The device under
test read the value of resistance selected on the tester. If it does not, the device under test
may need calibration, adjustment, or repair.