ERROR MESSAGES
A. SBS-PT-200S; SBS-PT-100S; SBS-PT-200SB; SBS-PT-100SB1:
• "Err-0": overload of weighing platform
• "Err-5": reading is not stable
• "Err-C": calibration error - a weight with the correct defect was not placed on the
platform
• "
": low level of battery charge
B. SBS-PT-200SK; SBS-PT-100SK:
• "
": low level of battery charge
• "Err-0": overload of weighing platform
• "Err-Z": range of zero values does not agree when the unit is turned on
• "Err-S": reading is not stable
• "Err-C": calibration error - a weight with the correct defect was not placed on the
platform
C. SBS-PT-60SNL; SBS-PT-150SNL:
• "Err-0": overload of weighing platform
• "Err-2": the range of zero values does not agree when the unit is turned on
• "Err-5": reading is not stable
• "Err-C": calibration error - a weight with the correct defect was not placed on the
platform
• "
": low level of battery charge
D. SBS-PT-30SN; SBS-PT-150SN:
• "Err-0": overload of weighing platform
• "
": low level of battery charge
E. SBS-LW-40:
• "Err-0": overload of weighing platform
• "LoB": low battery level
3.4. Cleaning and maintenance
a)
Use only non-corrosive cleaning agents for cleaning the surfaces.