Descargar Imprimir esta página

Aeroneb Drive Go 7090 Manual De Instrucciones página 26

Publicidad

Idiomas disponibles

Idiomas disponibles

Recommended separation distances between portable and mobile RF
communication equipment and the Aeroneb Go
The Aeroneb Go nebulizer system is intended for use in the electromagnetic
environment in which radiated RF disturbances are controlled. The customer
or the user of the Aeroneb Go can help prevent electromagnetic interference
by maintaining a minimum distance between port and mobile RF
communications equipment (transmitters) and the Aeroneb Go nebulizer
system as recommended below, according to the maximum output power
of the communications equipment.
Rated maximum
output power of
transmitter
W
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the
recommended separation distance d in metres (m) can be estimated using the
equation applicable to the frequency of the transmitter, where P is the maximum
output power rating of the transmitter in watts (w) according to the transmitter
manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher
frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures, objects
and people.
Separation distance according to frequency of
150 kHz to
80 MHz
d = [1.17] √P
0.12
0.37
1.17
3.70
11.70
transmitter
m
80 MHz to
800 MHz
d = [1.17] √P
0.12
0.37
1.17
3.70
11.70
800 MHz to
2.5 GHz
d = [2.33] √P
0.23
0.75
2.33
7.36
23.30