Immunity test
Power
frequency
(50/60 Hz)
Magnetic field
IEC 61000-4-8
Note: Ut is the a.c. mains voltage prior to application of the test level.
Guidance and manufacturer's declaration – electromagnetic immunity
The Aeroneb Go nebulizer system is intended for use in the electromagnetic
environment specified below. The customer or the user of the Aeroneb Go
should assure that it is used in such an environment.
Immunity
IEC/EN 60601
test
Test level
Conducted
3 Vrms
RF
IEC
150 kHz to
61000-4-6
80 MHz
Radiated RF
3 Vrms
IEC
80 MHz to
61000-4-3
2.5 GHz
IEC/EN 60601
Test level
3 A/m
Compliance
level
[3]V
[3]V
Compliance level
3 A/m
Electromagnetic environment –
Portable and mobile RF
communications equipment
should be used no closer to any
part of the Aeroneb Go, including
cables, than the recommended
separation distance calculated
from the equation applicable to
the frequency of the transmitter.
Recommended separation
distance
d = [1.17] √P
d = [1.17] √P... 80MHz to 800MHz
d = [2.33] √P... 800MHz to 2.5GHz
where P is the maximum output
power rating of the transmitter
in Watts (W) according to the
transmitter manufacturer and d
is the recommended separation
distance in metres (m).
Electromagnetic
environment –
guidance
Power frequency
magnetic fields
should be at levels
characteristic of
a typical location
in a typical
commercial
or hospital
environment.
guidance