• Remove both parts of the protective film from the adhesive pad.
5
• Pull the pretensioning element upwards, away from the base, as far as it will go.
W
PRECAUTION
Risk of injury
If the safety catch is in the
could be released accidentally.
Do not slide the safety catch to the
skin.
6
• Place the insertion device onto the chosen infusion site, holding it horizontally. It must be
positioned in such a way that the whole surface of each contact area on the base of the
insertion device is resting on the skin.
Tip: You should be able to comfortably reach the side of the head set which had the small tab on
the protective film. That is where you will later connect the transfer set.
• Slide the safety catch to the
7
• Press the release button. The cannula is inserted under the skin.
• Remove the insertion device.
If the head set of the infusion set is still in the insertion device, push the pretensioning element
back into the base. Continue with step
If the cannula is not properly positioned under the skin, remove the adhesive pad from the skin
position, the insertion device is no longer locked. The head set
position, until you have the insertion device placed on your
position. The insertion device is now unlocked.