Startup and Operation of the 4UHV Ion Pump
Controller from the Front Panel (LOCAL Mode)
To switch on the High Voltage (HV), the interlock of the HV cable (cable interlock)
NOTE
must be closed (connector inserted).
Proceed as follows to power the controller and apply voltage to the high voltage
(HV) connectors:
Connect the power cord to the rear panel
Switch off the controller.
Connect the HV cable and the interlock of the HV cable in the related
connectors on the rear panel.
Verify that the LOCAL mode is enabled, by pressing "MENU" then "MODE" push
button in sequence.
Hold down the HV ON/OFF button (2) while pressing the HV-x (3,4,5,6) button
where x is the number of the channel to be switched on/off.