Descargar Imprimir esta página

Tesmed max 830 Guía Rápida De Usuario página 8

Ocultar thumbs Ver también para max 830:

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 18
Intensity control:
Set initially a low level and then gradually increase. The intensity level
must be such that the treatment is pleasant.
Symbol
appears on the display
keys
and
, (from 1 to 99). The intensity rises or falls on all the selected channels at
the same time.
When you set the intensity, the clock begins to flash indicating that the equipment is
providing energy
(fig.
G).
ATTENTION: Programs from 21 to 99 are preset with different stimulation steps.
Each phase alternates moments of activity with moments of rest. It is not possible
to increase the intensity during the moment of rest. So do not increase the intensity
in the rest phases during program execution, but only during active
phases.
If you keep a low intensity, for example at level 1 or level 2, at the phase transition
during execution,the program may be reset. In this case, restart with the program
and keep the intensity a bit higher.
You can, however, access each channel during use to control the intensity independently:
simply press key
to select the channel whose intensity you want to change and then
use keys
and
, to increase or decrease the value.
values for the 4 channels can be different. At the end of the treatment, the Tesmed clears
the intensity setting for all output channels; you can now switch the equipment off by
pressing key
für 2 Sekunden gedrückt halten, oder neue Parameter einstellen, um
eine neue Behandlung zu beginnen. Die Erhöhung oder Senkung eines beliebigen
Parameters erfolgt mithilfe der Tasten
parameter, each time these keys are pressed, the parameter is increased or decreased
in steps of one unit, keep the keys pressed for faster increase/decrease of the parameter.
Error warning (electrodes disconnected)
In the case of a detached electrode, the device goes into error, displaying the word "Er",
flashing for about 5 ", in the channel where the problem occurred
All pulses are stopped immediately and "00" appears as the insensity value on the
display. The error message can mean that:
• There are no electrodes connected to the channel in error mode.
• The electrodes are not properly adhered to the skin: reposition the electrodes or
replace them if they are too old.
• The cable used to connect the electrodes is faulty: check by inverting two cables.
The system automatically resets itself once the intensity value increases.
"Keyboard block" function
Keeping the key
pressed, press key
the display
(fig.
L). Follow the same procedure to unlock the keyboard.
"ON block" function
Useful to avoid a random switch on.
With the device on, keeping the
the symbol
and the Tesmed logo will appear on the display
same procedure to switch the equipment on.
99
99
levels
(Fig. F)
and you can now adjust the intensity using
and
, to increase or decrease the value of any
and release it. Symbol
and
keys pressed simultaneously for 3 seconds,
Fig. H
shows that the intensity
(fig.
I).
will appear on
(Fig.
M). Follow the
8

Publicidad

loading

Productos relacionados para Tesmed max 830

Este manual también es adecuado para:

Max 5Max 7.8