1. Remove the lid and fill it with the liquid from the tube. Any froth should
remain in the tube.
2. Dip the head of the Reveal 3-D device into the liquid in the
lid. Ensure the cavity is saturated with the liquid.
3. Leave the cavity saturated until liquid is seen running in the
test window.
4. Place device on a flat surface and allow test to develop for
5 minutes.
NOTE: It is essential to place the device flat on a level surface
as soon as the liquid has entered the test window to stimulate
flow through the device. Additionally, the devices are pre-striped
with pale green loading dye in positions T (test), O (overload), and
C (control). The loading dye is removed from the test window as the sample flows through the device.
Read the result 5 minutes after dipping. Observations after 6 minutes may be inaccurate due to
overdevelopment of the device.
Negative result
No line at position T (test): Level of coconut below detection limit.
(See limitations section).
Positive result
Any intensity of line at position T (test): Level of coconut above
detection limit.
High results
No line is visible at position O (overload) and a line is faintly visible or absent at position T: Sample
is overloaded with coconut.
Invalid results
If no line appears at position C (control), then the test is invalid.
Questions? Call Neogen at 800/234-5333 or 517/372-9200
Questions? Call 800/234-5333 or 517/372-9200
SAMPLE TESTING
READING RESULTS
High
Negative
Positive
High