HBP-1320-E main.book Page 19 Tuesday, May 8, 2018 5:16 PM
Cuff Selection and Connection
Selecting the cuff
Warning
If a cuff is used on a patient with an infection, treat the cuff as medical waste, or disinfect before
reuse.
Caution
• Do not use the cuff if it is damaged or has holes.
• Use the appropriate cuff size to ensure correct measurements. If a cuff that is too large is used, the
measured blood pressure value tends to be lower than the actual blood pressure. If a cuff that is too
small is used, the measured blood pressure value tends to be higher.
Note:
It is important to use the correct sized cuff for a patient in order to get an accurate reading.
Measure the circumference of the patient's arm and select the cuff size
that is appropriate for the circumference.
Select the cuff that is suitable for the patient from the cuffs below.
Cuff name
GS CUFF2 XL*
GS CUFF2 L
GS CUFF2 M
GS CUFF2 S*
GS CUFF2 SS*
* Available as an optional accessory.
Connecting the cuff
Connect the cuff tube to the NIBP connector on the unit.
Caution
Only OMRON GS CUFF2 can be used with this device. The use of any other cuff may result in incorrect
measurement.
Note:
• Make sure the connections are tight.
Arm circumference
(cm)
(inch)
42 - 50
17 - 20
32 - 42
13 - 17
22 - 32
9 - 13
17 - 22
7 - 9
12 - 18
5 - 7
EN