HBP-1320-E main.book Page 24 Tuesday, May 8, 2018 5:16 PM
Maintenance
Maintenance Inspection and Safety Management
The HBP-1320 must be maintained to ensure functionality and to secure the safety of patients and operators.
Daily checks and maintenance should be performed by the operator. (page 25)
In addition, qualified personnel are necessary to maintain the performance and the safety, and to conduct
periodic inspections. We recommend that the verification test be performed at least once a year.
Example of connection for pressure accuracy confirmation:
1. Display the pressure accuracy confirmation screen as explained in
"Utility Mode" on page 17.
Display "0" in the pressure accuracy confirmation screen.
2. Connect the blood pressure monitor, the calibrated reference pressure
gauge, and the cuff and inflation bulb.
3. Check the pressure value of the blood pressure monitor and the
pressure value of the calibrated reference pressure gauge.
Note:
• Make sure that the blood pressure monitor reading is within ±3 mmHg compared to the calibrated
reference pressure gauge.
• To rapidly deflate the cuff, press the [
To repeat "Pressure Accuracy Confirmation", turn off the power and repeat the procedure from
step 1 in "Utility Mode" on page 17.
Cleaning of the Device
Cleaning and disinfecting should be performed in accordance with your facility's infection control
practice.
Surface cleaning
Wipe with a cloth that has been moistened with isopropyl alcohol diluted to 50 v/v%, or ethyl alcohol
(disinfection alcohol) diluted to 80 v/v% or less and wrung out.
Do not wipe the DC jack or allow it to become wet.
Removing dust
Use a moistened cotton bud to remove dust that has accumulated on the vent ports.
Service
The device requires no routine service other than cleaning, and visually checking the cuffs, tubing, etc.
Caution
• Do not sterilize by autoclave or gas sterilization (EOG, formaldehyde gas, high-concentration ozone, etc.).
• If using an antiseptic solution for cleaning, follow the instructions of the manufacturer.
Cuff Care
Wipe clean on the surface of the cuff with a cloth moistened with a 70 v/v% dilution of isopropyl alcohol,
or a 80 v/v% or less dilution of disinfection ethanol (ethyl alcohol).
Do not allow any liquids inside the cuff. If a liquid gets in the cuff, dry the inside well.
24
Display the pressure value.
] button.